Mouse IgG1 fluorescence controls are used to estimate the amount of nonspecific binding, as well as binding mediated by interaction with Fc receptors, using a BD FACS™ brand flow cytometer.
The flow cytometer must be equipped to detect light scatter and the appropriate fluorescence, and be equipped with appropriate analysis software (such as BD CellQuest™ or BD LYSYS™ II software) for data acquisition and analysis. Refer to your instrument user’s guide for instructions.
APC-Cy™7 is a tandem fluorochrome that combines APC and a cyanine dye (Cy7). Special precautions must be taken with APC-Cy7 conjugates, and cells stained with them, to protect the fluorochrome from long-term exposure to light. Some APC-Cy7 conjugates show changes in their emission spectra with prolonged exposure to paraformaldehyde. Fixed cells should be analyzed within 4 hours of fixation in paraformaldehyde or transferred to a paraformaldehyde-free buffer for overnight storage. Due to nearly identical excitation and emission properties but different spillover characteristics, APC-Cy7 and APC-H7 cannot be used simultaneously.
Resources & Tools | ||||||
---|---|---|---|---|---|---|
SpectrumViewer | Download MSDS |
The antibody reagent is stable until the expiration date shown on the label when stored at 2° to 8°C. Do not use after the expiration date. Do not freeze the reagent or expose it to direct light during storage or incubation with cells. Keep the outside of the reagent vial dry.
Do not use the reagent if you observe any change in appearance. Precipitation or discoloration indicates instability or deterioration.