CD19 is intended for in vitro diagnostic use in the identification of cells expressing CD19 antigen, using a BD FACS™ brand flow cytometer.
The flow cytometer must be equipped to detect light scatter and the appropriate fluorescence, and be equipped with appropriate analysis software (such as BD CellQuest™ or BD LYSYS™ II software) for data acquisition and analysis. Refer to your instrument user’s guide for instructions.
FITC, fluorescein isothiocyanate, is a fluorochrome with a molecular weight of 389 Da. FITC is sensitive to pH changes and photobleaching. Due to nearly identical excitation and emission properties but different spillover characteristics, FITC and Alexa Fluor® 488 cannot be used simultaneously. FITC is relatively dim and should be reserved for highly expressed markers whenever possible.
Resources & Tools | ||||||
---|---|---|---|---|---|---|
Spectrum Viewer | Panel Designer | SpectrumViewer | Download MSDS | CFDA License | IFU |
The antibody reagent is stable until the expiration date shown on the label when stored at 2° to 8°C. Do not use after the expiration date. Do not freeze the reagent or expose it to direct light during storage or incubation with cells. Keep the outside of the reagent vial dry.
Do not use the reagent if you observe any change in appearance. Precipitation or discoloration indicates instability or deterioration.
BD是世界上最大的生产和销售医疗设备、医疗系统和试剂的医疗技术公司之一。致力于提高全世界人类的健康水平。BD专注于改进药物治疗,提高传染性疾病诊断的质量和速度,推进新型药物和疫苗的研究与发现。公司于1897年在纽约成立,总部位于美国新泽西州的富兰克林湖,业务可分为BD医疗、BD诊断、BD生物科学三大类,生产销售包括医用耗材、实验室仪器、抗体、试剂、诊断等产品。